+1 (314) 370-6046 oderKontakt

Schwein MMP2 Kit Elisa (A5705)

Versandkosten
Vorlaufzeit
Lieferung in 7-10 Werktagen
Telefon
+1 (314) 370-6046
Mo bis Frei, 8am - 4pm AST
Email
orders@antibodies.com
Produktname
Porcine MMP2 ELISA Kit
Beschreibung
Porcine MMP2 ELISA Kit is a sandwich Enzyme-Linked Immunosorbent Assay (sELISA) designed for the in vitro quantitative determination of porcine MMP2 in serum, plasma, tissue homogenates, cell lysates or other biological fluids.
Assay-Typ
Sandwich (quantitative)
Assay-Prinzip
Porcine MMP2 ELISA Kit (A5705) employs the sandwich enzyme immunoassay technique for the quantitative measurement of porcine MMP2 in serum, plasma, tissue homogenates, cell lysates or other biological fluids. An antibody specific for MMP2 has been pre-coated onto a 96-well microtiter plate. The standards and test samples are added into the wells and the MMP2 present in each sample is bound to the wells by the immobilized antibody. Following incubation, the wells are washed and then incubated with Biotinylated Anti-MMP2 Antibody, which binds the captured MMP2 present in each well. Following incubation, unbound biotinylated detection antibody is removed by washing, and an HRP-Avidin conjugate is added to the wells and the microtiter plate is incubated. Following incubation and washing, TMB substrate solution is then used to visualize the HRP enzymatic reaction by catalysis to produce a blue-coloured product that changes to yellow after addition of acidic stop solution. The density of yellow is proportional to the amount of MMP2 captured in each well. The concentration of MMP2 can then be calculated by reading the O.D. absorbance at 450nm in a microplate reader and referring to the standard curve.
Plattform
Pre-coated Microplate (12 x 8 well strips)
Nachweismethode
Colourmetric
Instrument
Colorimetric Microplate Reader
Probentyp
Serum, plasma, tissue homogenates, cell lysates or other biological fluids.
Sensitivität
33 pg/ml
Detektionsbereich
78.1-5000 pg/ml
Testdauer
4h 30m
Reaktivität
Porcine
Recovery
Sample Type n Range Average
Serum 5 80% - 102% 91%
EDTA Plasma 5 81% - 100% 90%
Heparin Plasma 5 80% - 89% 84%
Linearity
Sample Type 1:2 1:4 1:8 1:16
Serum (n=5) 87-91% 87-107% 74-101% 92-97%
EDTA Plasma (n=5) 90-105% 84-101% 90-101% 79-108%
Heparin Plasma (n=5) 84-95% 92-105% 82-105% 89-91%
Precision
Intra-assay CV: < 8%, Inter-assay CV: < 10%
Allgemeine Hinweise
Information online is representative. Always refer to the Product Manual inside the kit.
Lagerung
Store kit components at +4°C or -20°C as instructed. Do not use past expiration date!
Synonyms
72 kDa gelatinase, 72 kDa type IV collagenase, 72kD type IV collagenase, CLG 4, CLG 4A , CLG4, CLG4A, Collagenase Type 4 alpha, Collagenase type IV A, EC 3.4.24.24, Gelatinase A, Gelatinase alpha , Gelatinase neutrophil , Matrix metallopeptidase 2 (gelatinase A, 72kDa gelatinase, 72kDa type IV collagenase), Matrix metallopeptidase 2 gelatinase A 72kDa gelatinase 72kDa type IV collagenase, Matrix metalloproteinase 2, Matrix metalloproteinase 2 (gelatinase A, 72kD gelatinase, 72kD type IV collagenase), Matrix metalloproteinase 2 (gelatinase A, 72kDa gelatinase, 72kDa type IV collagenase), Matrix metalloproteinase II, Matrix metalloproteinase-2, MMP 2, MMP II, MMP-2, MMP2_HUMAN, MONA, Neutrophil gelatinase, PEX, TBE 1, TBE-1
Haftungsausschluss
Dieses Produkt ist nur für Forschungszwecke bestimmt. Es ist nicht für diagnostische oder therapeutische Zwecke bestimmt.
Kit Components
Item Quantity Storage
Pre-Coated 96 Well Microplate 12 x 8 Well Strips -20°C
Lyopholized Standard 2 Vials -20°C
Detection Solution A 120μl -20°C
Detection Solution B 120µl -20°C
Wash Buffer (30X) 20ml +4°C
Sample Dilution Buffer 45ml -20°C
TMB Substrate 9ml +4°C
Stop Solution 6ml +4°C
Plate Sealers 5 Adhesive Strips -
Required Materials
The following materials are not included in the kit, but are required to run this assay:
  • Microplate reader capable of measuring absorbance at 450nm ± 10nm.
  • Squirt bottle, multichannel pipette reservoir, or automated microplate washer.
  • Graph paper or computer software capable of generating logarithmic functions.
  • Test tubes and Eppendorf tubes to prepare standards and samples.
  • Multi- and single-channel pipettes and sterile pipette tips.
  • Deionized or distilled water.
  • Absorbent paper towels.
  • 37°C incubator.
Important Notes
  • Sample concentrations should be estimated before being used in the assay and a proper dilution factor should be selected to make the diluted target protein concentration fall in the optimal detection range of this kit.
  • Hemolysis can greatly impact the validity of test results. Take care to minimize hemolysis.
  • Samples should be brought to room temperature before starting the assay.
Bring all reagents and samples to room temperature before use.
We recommend that all standards and samples be assayed in duplicate.
Step 1
Prepare all reagents, samples, and working standards as directed in the previous sections.
Step 2
Determine the number of microplate strips required for the assay and insert them into the plate frame. Unused strips should be stored in a foil pouch at -20°C.
Step 3
Add 100μl of standard solutions to the standard wells. Note: A plate layout is provided to record standards and samples assayed.
Step 4
Add 100μl of samples to the sample wells.
Step 5
Cover with an adhesive plate sealer provided and incubate at 37°C for 2 hours.
Step 6
Remove the plate sealer and aspirate the liquid from the plate.
Step 7
Add 100μl of working Detection Solution A to each well.
Step 8
Cover with a new adhesive plate sealer and incubate at 37°C for 60 minutes.
Step 9
Remove the plate sealer, aspirate the liquid from the plate, and wash the plate three times with Wash Buffer. Soak wells with 300μl of Wash Buffer for 1-2 minutes each time. Note: Complete removal of liquid at each step is essential to good performance. After the last wash, remove any remaining Wash Buffer by aspirating or decanting. Blot the plate onto paper towels or other absorbent material. Do not let the wells dry completely at any time!
Step 10
Add 100μl of working Detection Solution B to each well.
Step 11
Cover with a new adhesive plate sealer and incubate at 37°C for 60 minutes.
Step 12
Remove the plate sealer, aspirate the liquid from the plate, and wash the plate five times with Wash Buffer. Soak wells with 300μl of Wash Buffer for 1-2 minutes each time.
Step 13
Add 90μl of TMB Substrate to each well. Cover with a new plate sealer and incubate at 37°C in the dark for 15-25 minutes (do not exceed 30 minutes). Note: This incubation time is for reference only, the optimal reaction time should be determined by the end-user and can be shortened or extended according to the actual color change. The reaction may be terminated when a gradient is apparent in the standard wells.
Step 14
Add 50μl of Stop Solution to each well. The color in the wells should change from blue to yellow immediately. Note: If the color in the wells is green or the color change does not appear uniform, gently tap the plate to ensure thorough mixing.
Step 15
Determine the optical density (O.D. value) of each well immediately using a microplate reader set to 450 nm.

Alternative Produkte zum Schwein MMP2 Kit Elisa (A5705)