+1 (314) 370-6046 oderKontakt

Human LDL Receptor Kit Elisa (A2141)

$585
Versandkosten
Vorlaufzeit
Lieferung in 7-10 Werktagen
Telefon
+1 (314) 370-6046
Mo bis Frei, 8am - 4pm AST
Email
orders@antibodies.com
Produktname
Human LDL Receptor ELISA Kit
Beschreibung
Human LDL Receptor ELISA Kit is a sandwich (quantitative) Enzyme-Linked Immunosorbent Assay (sELISA) designed for the in vitro quantitative determination of human LDL Receptor in tissue homogenates, cell lysates or other biological fluids.
Assay-Typ
Sandwich (quantitative)
Assay-Prinzip
Human LDL Receptor ELISA Kit employs the Sandwich Enzyme Immunoassay (ELISA) technique for the quantitative measurement of human LDL Receptor in human. An antibody specific for LDL Receptor has been pre-coated onto a 96-well microtiter plate. The standards and test samples are added into the wells and the LDL Receptor present in each sample is bound to the wells by the immobilized antibody. Following incubation, the wells are washed and then incubated with Biotinylated Anti-LDL Receptor Antibody, which binds the captured LDL Receptor present in each well. Following incubation, unbound biotinylated detection antibody is removed by washing, and Avidin conjugated to Horseradish Peroxidase (Avidin-HRP) is added to the wells and the microtiter plate is incubated. Following incubation and washing, TMB substrate solution is used to visualize the HRP enzymatic reaction by catalysis to produce a blue-coloured product that changes to yellow after addition of acidic stop solution. The density of yellow is proportional to the amount of LDL Receptor captured in each well. The concentration of LDL Receptor can then be calculated by reading the O.D. absorbance at 450nm in a microplate reader and referring to the standard curve.
Plattform
Pre-coated Microplate (12 x 8 well strips)
Nachweismethode
Colourmetric
Instrument
Colorimetric Microplate Reader
Probentyp
Tissue homogenates, cell lysates or other biological fluids.
Sensitivität
0.118 ng/ml
Detektionsbereich
0.312-20 ng/ml
Testdauer
1 - 4.5 Hours
Reaktivität
Human
Recovery
Sample Type n Range Average
Serum 5 80% - 102% 91%
EDTA Plasma 5 81% - 100% 90%
Heparin Plasma 5 80% - 89% 84%
Linearity
Sample Type 1:2 1:4 1:8 1:16
Serum (n=5) 87-91% 87-107% 74-101% 92-97%
EDTA Plasma (n=5) 90-105% 84-101% 90-101% 79-108%
Heparin Plasma (n=5) 84-95% 92-105% 82-105% 89-91%
Precision
Intra-assay CV: < 8%, Inter-assay CV: < 10%
Allgemeine Hinweise
Information online is representative. Always refer to the Product Manual inside the kit.
Lagerung
Store at +4°C. Do not use past expiration date!
Synonyms
FH, FHC, LDL R, LDLCQ2, Ldlr, LDLR_HUMAN, Low Density Lipoprotein Receptor, Low density lipoprotein receptor class A domain containing protein 3, Low density lipoprotein receptor familial hypercholesterolemia, Low-density lipoprotein receptor
Haftungsausschluss
Dieses Produkt ist nur für Forschungszwecke bestimmt. Es ist nicht für diagnostische oder therapeutische Zwecke bestimmt.

Alternative Produkte zum Human LDL Receptor Kit Elisa (A2141)

Oben