Principle of Assay
Mouse Follistatin ELISA Kit (A1926) employs the competitive enzyme immunoassay technique for the quantitative measurement of mouse Follistatin in serum, plasma or other biological fluids. The 96-well microtiter plate has been pre-coated with Follistatin antigen. During the incubation, Follistatin present in the samples or standards competes with the fixed amount of immobilized Follistatin for binding sites on the Biotinylated Anti-Follistatin Antibody. The more Follistatin present in a sample or standard, the less Biotinylated Anti-Follistatin Antibody that binds to the plate. Following incubation, unbound Biotinylated Anti-Follistatin Antibody is removed by washing, and an HRP-Avidin conjugate is added to the wells and the microtiter plate is incubated. Following incubation and washing, TMB substrate solution is then used to visualize the HRP enzymatic reaction by catalysis to produce a blue-coloured product that changes to yellow after addition of acidic stop solution. The density of yellow is inversely proportional to the amount of Follistatin present in each sample or standard. The concentration of Follistatin can then be calculated by reading the O.D. absorbance at 450nm in a microplate reader and referring to the standard curve.