Principle of Assay
Human CCL17 ELISA Kit is based on the quantitative sandwich enzyme-linked immunosorbent assay technique to measure concentration of human CCL17 in the samples. A monoclonal antibody specific for human CCL17 has been immobilized onto microwells. Standard or samples are pipetted into the wells, followed by the addition of biotin-linked detect antibody specific for CCL17, and CCL17 present is bound by the immobilized antibody and detect antibody following the first incubation. After removal of any unbound substances, streptavidin-HRP is added for a second incubation. After washing, substrate solution reacts with HRP and color develops in proportion to the amount of CCL17 bound by the immobilized antibody. The color development is stopped by addition of acid and the optical density value is measured by microplate reader.